Setting Up Your Oscilloscope for EFT Testing

This article explains how to properly configure an oscilloscope for EFT testing, covering sample rates, parameter acceptance, EMC pulse measurements, energy calculations, and efficient burst acquisition techniques.

Setting Up Your Oscilloscope for EFT Testing

A third type of electromagnetic compatibility (EMC) testing deals with how devices respond to electrical fast transients (EFT). EFTs are a series of fast, high frequency pulses, often occurring in bursts. These transient events are the result of electrical arcing. EFT pulse bursts occur when a power connection is made or broken, equipment is powered down or circuit breakers are switched. They also occur when inductive loads such as relays, switch contactors or heavy-duty motors produce bursts of narrow high-frequency transients on the power distribution system when de-energized.

Figure 1: The typical EFT test signal consists of multiple exponential pulses arranged as pulse bursts.

The typical compound waveform used for EFT testing is shown in Figure 1.

The pulse waveform for EFT testing is defined by a risetime of 5 ns and a pulse width of 50 ns. These pulses are combined into bursts of 5 to 50 pulses spaced at from 10 to 100 µs (10 to 100 kHz). The bursts are typically spaced as much as 300 ms apart. Figure 2 shows an EFT test signal, with two EFT bursts captured by an oscilloscope on channel 2 sampled at 1.25 GS/s.

The two EFT pulse burst are shown in the upper left-hand trace (the two pink blocks). A series of zoom-on-zoom traces are opened to show several timing epochs. Trace Z2 (second from top left) shows a single burst. The zooms keep expanding the horizontal scale until finally at Z8 (bottom right) we see a single EFT pulse.

Following are six, important things to do to make sure you get the best EFT test measurements from your oscilloscope.

1 - Set a Sufficiently High Sample Rate

Given the fast rise time of the EFT pulse, the oscilloscope’s sample rate should be high enough to characterize the fast edge and to guarantee that the peaks are adequately sampled. For the examples shown here, the sample rate was set to 1.25 GS/s. This assures over six samples on the fast edge and a good characterization of the peaks of the waveform.

2 - Use Parameter Acceptance to Isolate Pulse Repetition Rate

One measurement of interest is the pulse repetition rate within the burst. The frequency parameter can be used to make this measurement, but because the source waveform (C2) is a burst waveform, it will show the pulse repetition rate along with the burst repetition rate. To look at only the higher frequency pulse repetition rate, use parameter acceptance to report only measured values that fall within a user-defined range. Figure 2 shows the Accept subdialog set up for a range of 9 to 11 kHz. Parameter P1 shows the corrected measurement of 10 kHz.

Figure 2: Acquisition of two EFT bursts at 1.25 GS/s, zoomed to show several timing epochs.

3 - Use Parameter Acceptance to Measure Inter-burst Gap Time

It is often asked, "how do you measure the gap time in between the bursts of EFT pulses?" You could simply measure negative pulse width (between a falling and a rising edge), but then you'd get the negative pulse width between each of those individual EFT pulses, not just the inter-burst gaps. Again, this is a case for using parameter acceptance, as shown in parameter P2 of Figure 3. Set up a pulse Width @ Level parameter with a negative polarity, then accept the lowest and highest measurement values that potentially define the inter-burst gaps. In this example, the limits of 80 milliseconds and 1 second exclude those individual high-frequency pulse widths, only measuring the gap time in between EFT bursts.

Figure 3: Parameter acceptance reports only the negative widths that are between 80 ms and 1 s to capture the gap between pulse bursts.

4 - Set Measurement Thresholds Correctly

Other typical EFT measurement are the rise time and width of the individual pulses. As with other types of EMC test waveforms that are not bi-level and therefore do not have a top and base, non-rectangular EFT pulses require that measurements be made using the % 0 - Max amplitude reference level, as shown in Figure 4.

Figure 4: Measuring EFT pulse rise time requires setting the measurement threshold based on the 0 to maximum amplitude levels, not top and base.

5 - Use Math on Parameters to Calculate Energy Correctly

Some EFT standards require the measurement of pulse maximum amplitude, pulse area and energy.

Figure 5: Measuring peak (maximum) amplitude, area under the pulse, and pulse energy using waveform math and parameter math.

To calculate energy, follow these steps (Figure 5):

  1. Measure maximum pulse amplitude (P5).
  2. Use a Math function to square the pulse voltage (F2), taking the product of the input channel times the input channel (C2*C2).
  3. Integrate the area under the squared voltage by applying the Area measurement (P6) to the squared voltage math function (F2).
  4. Use Math on Parameters to divide the area measurement by the oscilloscope’s input impedance, in this case 50 Ω:
    1. Set a constant of 50 Ω using the P Const operator (P7).
    2. Take the ratio of the area (P6) and the constant (P7) using the P Ratio operator (P8).

The energy of our EFT pulse is automatically calculated at 7.35 micro-Joules.

6 - Use Sequence Mode to Acquire Only Pulses of Interest

The waveforms acquired for EFT measurements can have very small duty cycles, being short pulses separated by long intervals with no activity. This type of waveform can use up a lot of oscilloscope memory. One way to extend memory is to use sequence mode acquisition. Sequence mode captures smaller segments of interest (of a user-defined length) based on a triggering event. Once triggered, the oscilloscope only acquires data for the duration of the interesting activity, storing one segment in memory, then stops acquiring until the next trigger. Figure 6 shows an EFT acquisition made for ISO 7637 that uses sequence mode.

Figure 6: 20 pulses captured using sequence mode at 50 MS/s using only 5000 samples per segment. The entire acquisition uses only 100,000 samples. Time stamps show the time between pulses.

The event of interest is a narrow pulse (15 µs width), followed by an electrical idle for the balance of the 500 ms period, representing a duty cycle of only 0.003%. By using sequence mode with a trigger set on the 15 µs width, you can capture 20 segments at 50 MS/s using only 5000 samples per segment, a total of only 100,000 samples, instead of the 50 million samples it would take to capture 10 such pulses using real-time sampling mode.


Mike Hertz presented a 2-part webinar series How to Get the Most Out of Your EMC/EMI Lab Oscilloscope. Part 1: How to Optimize Your Oscilloscope for Basic EMC/ESD Pulse Measurements of that webinar series covers more details on this topic. Part 2: Advanced EMC/EMI Pulse Measurement Techniques and Examples goes into more details on how to meet the exact needs of the IEEE and IEC standards for these measurements. Click the button below to watch Part 1, or download the transcript to preview the webinar content.

Great! You’ve successfully signed up.

Welcome back! You've successfully signed in.

You've successfully subscribed to Oscilloscope Blog | Teledyne LeCroy.

Success! Check your email for magic link to sign-in.

Success! Your billing info has been updated.

Your billing was not updated.